LIVANOVA DEUTSCHLAND GMBH HEATER COOLER 3T (230 VOLT); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 16-02-80 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/09/2024 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland received a report that, during priming, the circulation motor of a heater-coolersystem 3t was noisy and patient pump 1 stuck.Cleaning the motor and the pump did no solve the issue.There was no patient involvement.
|
|
Manufacturer Narrative
|
A.1.-a.5.There was no patient involvement.G.5.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in india.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
See intial report.
|
|
Manufacturer Narrative
|
The unit was installed in 2017 and the complaint database review revealed that a similar event was reported against the cardioplegia motor and was caused by a faulty circuit board.Based on the above facst the result of investigations performed on similar cases, the root cause of the event has been traced back to a failure of the stirring mechanism, which is likely to assume was stuck or no longer moving freely.Environmental conditions, such as high temperatures, humidity, condensation, and water infiltration may have contributed to the failure of the motor.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
|