DEPUY INTERNATIONAL LTD - 8010379 DEPUY ASR XL FEM IMP SIZE 55; ASR HIP SYSTEM : HIP METAL FEMORAL HEADS Back to Search Results
Catalog Number 999890155
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907)
Event Date 02/28/2012
Event Type  Injury  
Event Description
Metallosis and pseuodotumor.Replacement needed.Doi: (b)(6) 2007.Doe: (b)(4\6) 2012.
 
Manufacturer Narrative
Product complaint # (b)(4).No 510(k) number provided because this implant is sold internationally with different indications for use; it is currently sold in the us under a different part number.The correction/removal reporting number listed applies to the corresponding product code sold domestically.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint #
=
> (b)(4).Investigation summary
=
> an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : according to the information received, ¿metallosis and pseuodotumor.Replacement needed¿.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that light marks were observed all over the outer surface of the depuy asr xl fem imp, expected due to the metal-metal interaction for over 5 years.Additionally, signs of wear and what appears to be black organic material were found in the taper area, where the femoral stem is assembled.The observed condition of the device might contribute the reported metallosis event.It is worth mentioning, that the asr platform was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.With the information provided is not possible to determine a potential cause at this moment.The mode of failure of the device is multi-factorial and consideration must be given to all other potential influences such as surgical process, patient variables (e.G.Activity level and use), anatomical considerations and patient changes over time.Any conclusions from the investigational input provided have to be placed into context with all other relevant factors.A dimensional inspection and functional testing were not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the depuy asr xl fem imp would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : there is no evidence suggesting manufacturing or material error as a potential cause for the incident involving the asr platform.Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPUY ASR XL FEM IMP SIZE 55
Type of Device
ASR HIP SYSTEM : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18856476
MDR Text Key337112777
Report Number1818910-2024-05324
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeIT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/31/2020
Device Catalogue Number999890155
Device Lot Number1852476
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/27/2024
Initial Date FDA Received03/07/2024
Supplement Dates Manufacturer Received04/05/2024
05/20/2024
05/23/2024
Supplement Dates FDA Received04/08/2024
05/23/2024
05/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Treatment
ASR ACETABULAR IMPLANT 62.; UNK HIP FEMORAL HEAD METAL ASR.
Patient Outcome(s) Required Intervention;
-
-