This report is being supplemented to provide additional information, based on the legal manufacturer's final investigation and device evaluation.Additionally, to provide correction to the initial with information inadvertently left out (g2), to provide additional information received through follow up (b5) and to provide an update to fields (d8, d9, h4, and h4).The device was evaluated by olympus.And it was confirmed, that [alarm sound is generated and the front panel is turned off occurred.A review of the device history record found no deviations, that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely, the phenomenon "alarm sound is generated and the front panel is turned off" occurred, due to mother board failure.However, the root cause of the phenomenon could not be identified.Olympus will continue to monitor field performance for this device.
|
The device malfunction occurred in the middle of an unknown procedure.The device was reset and the procedure continued.A delay of approximately 30 seconds occurred, while the patient was under anesthesia.To reset the device and recycle the power.The procedure was completed using the same device.No error messages were displaced, due to the device malfunction.
|