• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP*MCA MED LONG APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP*MCA MED LONG APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number MCM30
Device Problem Failure to Form Staple (2579)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 3/7/2024.D4: batch # unk.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that during a mastectomy procedure, when attempting to clip the vessel the clips would either be too tight and almost severe the vessel or the clip would be too lose and not forming at all.A new device was used to complete the procedure.There were no adverse consequences for the patient.
 
Manufacturer Narrative
(b)(4).Date sent: 4/11/2024.D4: batch # x70525: investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the mcm30 device was received with no damage to the external components.In addition, the tyvek was returned along with the instrument.Upon cycling, the instrument was noted to be empty and locked out.The instrument is designed to lockout after all the clips have been fired; therefore a potential cause for the customer reported experience is the firing of all of the clips, as a result, the instrument could no longer be fire due to the activation of the lockout mechanism.As the device was returned empty for evaluation we are unable to investigate further the issue of ¿malformed clips".The device was disassembled to verify the condition of the internal components and no anomalies were noted. the event described could not be confirmed as the device was returned empty.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key18857584
MDR Text Key337745836
Report Number3005075853-2024-01992
Device Sequence Number1
Product Code GDO
UDI-Device Identifier10705036002482
UDI-Public10705036002482
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K820837
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMCM30
Device Lot Number779C17
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/21/2024
Initial Date FDA Received03/07/2024
Supplement Dates Manufacturer Received03/15/2024
Supplement Dates FDA Received04/11/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/21/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-