• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIVATORS INC ADVANTAGE PLUS ENDOSCOPE REPROCESSING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVATORS INC ADVANTAGE PLUS ENDOSCOPE REPROCESSING SYSTEM Back to Search Results
Device Problem Fire (1245)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2024
Event Type  malfunction  
Event Description
The user facility reported that the compressor to their advantage plus endoscope reprocessing system caught fire.No report of injury.
 
Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the unit and found that the compressor was damaged.The technician replaced the compressor, tested the unit, confirmed it to be operating according to specification and returned the unit to service.The compressor subject of the event was discarded and not returned for evaluation.As the device is not available for evaluation, a root cause of the reported event could not be determined.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTAGE PLUS ENDOSCOPE REPROCESSING SYSTEM
Type of Device
ENDOSCOPE REPROCESSING SYSTEM
Manufacturer (Section D)
MEDIVATORS INC
14605 28th avenue n
minneapolis MN 55447
Manufacturer (Section G)
MEDIVATORS INC
14605 28th avenue n
minneapolis MN 55447
Manufacturer Contact
daniel davy
14605 28th ave north
minneapolis, MN 55447
4403927453
MDR Report Key18861696
MDR Text Key337162415
Report Number2150060-2024-00015
Device Sequence Number1
Product Code FEB
UDI-Device Identifier00677964044755
UDI-Public00677964044755
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/09/2024
Initial Date FDA Received03/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/22/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-