FARAPULSE, INC. FARAWAVE PULSED FIELD ABLATION CATHETER, 31MM; CARDIAC IRREVERSIBLE ELECTROPORATION SYSTEM CATHETER
|
Back to Search Results |
|
Lot Number 0008021427 |
Device Problem
Positioning Failure (1158)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/20/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The farawave pulsed field ablation catheter was returned to boston scientific for analysis.Upon receipt at the post market quality assurance laboratory the catheter underwent visual inspection, x-ray inspection, and functional testing.No outer abnormalities were observed.The device was put on the x-ray to look for any potential abnormalities that could have contributed to deployment issues, but no anomalies were noted.A guidewire was inserted through the device and an attempt was made to deploy the catheter.While moving the slider switch, the spline cage remained undeployed.It was observed that the guidewire lumen was no longer adhered to the tip of the spline cage.The device was dissected to look for any abnormalities that could have contributed to the delamination of the guidewire lumen.Nothing out of the ordinary was noted.The reported clinical observations were confirmed.
|
|
Event Description
|
It was reported that during an ablation procedure to treat atrial fibrillation a farawave pulsed field ablation catheter was selected for use.The catheter was prepared properly before insertion and was able to deploy to the basket and flower configurations and then un-deploy.After the first ablation application in the basket configuration the catheter could not be deployed to the flower configuration despite the slider moving back.The catheter was undeployed, but then could not be deployed to either configuration.The catheter was removed from the patient and examined.No resistance was felt when moving the slider and there were no problems with the guidewire.The catheter was replaced and the procedure was completed successfully.No patient complications were reported.The device is expected to be returned for analysis.The farawave pulsed field ablation catheter was received and analyzed at boston scientific's post market laboratory.Upon receipt at our post market quality assurance laboratory it was found, while testing deployment, that the guidewire lumen was no longer adhered to the tip of the spline cage, resulting in the inability to deploy the catheter.
|
|
Manufacturer Narrative
|
The farawave pulsed field ablation catheter was returned to boston scientific for analysis.Upon receipt at the post market quality assurance laboratory the catheter underwent visual inspection, x-ray inspection, and functional testing.No outer abnormalities were observed.The device was put on the x-ray to look for any potential abnormalities that could have contributed to deployment issues, but no anomalies were noted.A guidewire was inserted through the device and an attempt was made to deploy the catheter.While moving the slider switch, the spline cage remained undeployed.It was observed that the guidewire lumen was no longer adhered to the tip of the spline cage.The device was dissected to look for any abnormalities that could have contributed to the delamination of the guidewire lumen.Nothing out of the ordinary was noted.The reported clinical observations were confirmed.
|
|
Event Description
|
Clinical study name: faradise, clinical study id: (b)(6).It was reported that the catheter could not be deployed properly after preparation.During an ablation procedure to treat atrial fibrillation a farawave pulsed field ablation catheter was selected for use.The catheter was prepared properly before insertion and was able to deploy to the basket and flower configurations and then un-deploy.After the first ablation application in the basket configuration the catheter could not be deployed to the flower configuration despite the slider moving back.The catheter was undeployed, but then could not be deployed to either configuration.The catheter was removed from the patient and examined.No resistance was felt when moving the slider and there were no problems with the guidewire.The catheter was replaced and the procedure was completed successfully.No patient complications were reported.The farawave pulsed field ablation catheter was received and analyzed at boston scientific's post market laboratory.Upon receipt at our post market quality assurance laboratory it was found, while testing deployment, that the guidewire lumen was no longer adhered to the tip of the spline cage, resulting in the inability to deploy the catheter.
|
|
Search Alerts/Recalls
|
|
|