ICU MEDICAL COSTA RICA LTD. PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 146870489
Device Problem Disconnection (1171)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/13/2024
Event Type  malfunction  
Manufacturer Narrative
The device is available to be returned for evaluation; however, it has not yet been received.
 
Event Description
The event involved a primary plum set, clave secondary port, clave y-site, secure lock, 103 inch where it was reported that for primary iv tubing line, when changing tubing, writer unscrewed tubing at hub of iv access.The screw at end of tubing fell apart in writer's hands and fell off tubing, leaving the writer the need to change entire tubing set as tubing is now unable to screw in and remain attached.This was a new iv tubing/unopened package.The product did not appear defective in packaging.The packaging was not saved so they were unable to determine lot number for this specific product.It was further reported that they were able to resolve the event.There was patient involvement, no human harm, and no medical intervention required.
 
Manufacturer Narrative
A picture was returned showing the end of two sets where the secure lock male adapter can be seen.The top set appeared to be missing the locking component.The complaint of the screw at the end of the tubing falling apart can be confirmed based on the returned image.Without the return of the used sample a comprehensive failure investigation cannot be performed, and a cause cannot be determined.A lot history review could not be conducted because no lot number(s) was/were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL COSTA RICA LTD.
zona franca global
la aurora heredia
CS 
Manufacturer Contact
michael visocnik
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18862931
MDR Text Key337187672
Report Number9615050-2024-00218
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number146870489
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/15/2024
Initial Date FDA Received03/07/2024
Supplement Dates Manufacturer Received03/27/2024
Supplement Dates FDA Received04/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-