• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 20 CM; CATHETER PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 20 CM; CATHETER PERCUTANEOUS Back to Search Results
Catalog Number CU-25122-F
Device Problem Unraveled Material (1664)
Patient Problem Insufficient Information (4580)
Event Date 01/31/2024
Event Type  malfunction  
Event Description
It was reported that "swg unraveled during removal." the device was removed and replaced.Another set was used and was successful.No paitent harm or injury.The current status of the patient is "unknown" at this time.
 
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that "swg unraveled during removal." the device was removed and replaced.Another set was used and was successful.No paitent harm or injury.The current status of the patient is "unknown" at this time.
 
Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.The instructions-for-use (ifu) provided with this kit warns the user, "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause component damage or breakage.If damage is suspected or withdrawal cannot be easily accomplished, radiographic visualization should be obtained, and further consultation requested." a device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 20 CM
Type of Device
CATHETER PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18865130
MDR Text Key337199899
Report Number3006425876-2024-00220
Device Sequence Number1
Product Code DQY
UDI-Device Identifier00801902100276
UDI-Public00801902100276
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K862056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberCU-25122-F
Device Lot Number71F23A2935
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/15/2024
Initial Date FDA Received03/08/2024
Supplement Dates Manufacturer Received04/02/2024
Supplement Dates FDA Received04/04/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NONE REPORTED; NONE REPORTED
-
-