RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 V60PLUS VENTILATOR |
Device Problem
Tidal Volume Fluctuations (1634)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/05/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
E: (b)(6) hospital.(b)(6).Phone: (b)(6).
|
|
Event Description
|
Philips received a complaint by the customer on the v60 indicating that the device is getting an error message indicating the tidal volume is too high.The device was in use on a patient at the time the reported issue was discovered; however, there was no harm to the patient or user.No medical intervention provided to the patient, nor a delay was noted.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed the reported problem.The rse recommended to replace the flow sensor assembly and dispatched a field service engineer for on-site repair.Investigation ongoing.
|
|
Manufacturer Narrative
|
H10: it was reported there was no patient involvement at the time the issue was discovered.The manufacturer's field service engineer (fse) evaluated the device, and the reported issue was not duplicated, no abnormality was confirmed.The device passed required performance verification tests per philips standards and was returned to service.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.Based on information provided and/or service performed it could not be confirmed unit did not meet product specifications.No fault found.The device was not being used for treatment when the reported event occurred.
|
|
Search Alerts/Recalls
|
|
|