Model Number V60 V60PLUS VENTILATOR
Device Problem Computer Software Problem (1112)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2024
Event Type  malfunction  
Event Description
Philips received a complaint from the customer, reporting that the v60 ventilator displayed an error code for watchdog test failed.There was no patient involvement when the issue occurred.No patient or user harm reported.A philips remote service engineer (rse) the customer had previously replaced the flow sensor assembly, but the device was now displaying the error code.The rse advised the customer to reinstall the old flow sensor assembly to see if it would reoccur.The rse explained that if it remained after installing the old flow sensor assembly, the central processing unit (cpu) would need to be replaced.Investigation ongoing.
 
Manufacturer Narrative
The customer reported that a new flow sensor assembly was tested, and the issue was resolved.The device was returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18865542
MDR Text Key337213602
Report Number2518422-2024-12795
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1138747
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/27/2024
Initial Date FDA Received03/08/2024
Supplement Dates Manufacturer Received03/14/2024
Supplement Dates FDA Received03/21/2024
Date Device Manufactured01/09/2021
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-