MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION VERCISE GEVIA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION VERCISE GEVIA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS Back to Search Results
Model Number DB-1200-S
Device Problem Overheating of Device (1437)
Patient Problem Burning Sensation (2146)
Event Date 12/19/2023
Event Type  Injury  
Event Description
It was reported that the deep brain stimulation (dbs) patient experienced burning sensations around the implantable pulse generator (ipg) site wherein causing the site to feel hot.This began occurring intermittently overtime after the patient had undergone a mammogram.A database analysis was performed and found no anomalies with the ipg.The patient underwent an exploratory surgery of the ipg site, and ipg extension connection sites and the physician assessed the ipg, and extension connections were fine.The extensions were reconnected to the ipg.The patient was doing well post-operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERCISE GEVIA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key18866489
MDR Text Key337226508
Report Number3006630150-2024-01339
Device Sequence Number1
Product Code NHL
UDI-Device Identifier08714729984443
UDI-Public08714729984443
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/18/2022
Device Model NumberDB-1200-S
Device Catalogue NumberDB-1200-S
Device Lot Number743853
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/14/2024
Initial Date FDA Received03/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/18/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age63 YR
Patient SexFemale
-
-