RESPIRONICS, INC. BIPAP A40 SYSTEM SILVER; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number R1111177 |
Device Problem
Infusion or Flow Problem (2964)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/19/2022 |
Event Type
malfunction
|
Event Description
|
The manufacturer received information alleging a ventilator's flow stopped temporarily.There was no harm or injury reported.During the evaluation of the device at the manufacturer's service center, the reported issue was unable to be confirmed.An error code was found in the ventilator's downloaded error log.The device's main board was replaced to resolve the issue.Overall check, cleaning, and functional testing was performed.
|
|