It was reported that the pressurewire x, wireless device calibrated and equalized successfully.The device was to be used in the diagonal branch lesion.However, measurement was unstable.The tip of the device was noted to be wavy and frayed.Therefore, the device was removed and pressurewire x, wireless device was used to complete the procedure.The transmitter status light was steady green.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
Visual and functional analysis was performed on the returned device.The reported pressure signal drift was not confirmed as the results of the functional testing were within the specifications for the pressurewire.The device was able to calibrate, equalize, and function as expected.The reported material frayed in the tip coil could not be confirmed.However, there was a stretch noted on the distal tip coil.The reported bend/kink/wavy in the tip coil was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no indication of a lot specific product quality issue.The investigation was unable to determine a cause for the reported pressure signal drift.It may be possible that procedural contaminants/contrast on the pressure sensor membrane or a system issue occurred contributing to the reported signal drift; however, this could not be confirmed.The reported wavy material was likely due to the circumstances of the procedure.It is likely that challenging anatomical conditions or excessive force inadvertently damaged the tip coil (such as bend, kink, or stretch) resulting in the reported issue.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.B3: date of event is estimated as (b)(6) 2024.
|