This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Corrected fields: h6 - medical device problem code is being corrected to "4048 - failure to clean adequately".A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 4 years since the subject device was manufactured.Based on the results of the investigation, the white foreign material identified was organic silicones and stearates, which emerged from the channel, including the auxiliary water channel.Organic silicones are components found in antifoam agents, drugs, etc., while stearates are material used in emulsifiers, cosmetics, etc.There was no damage to the area where the foreign material was detected, and it was unknown if reprocessing was performed according to the instructions for use.Therefore, the cause of the material remaining in the device could not be determined.The event can be detected by following the instructions for use which state: the operation manual describes how to inspect for the suggested events in ¿chapter 3 preparation and inspection, section 3.8 inspection of the endoscopic system¿ as below.[inspection of the auxiliary water feeding function].2 feed water from the syringe or the water tube.3 confirm that water is emitted from the auxiliary water channel at the distal end of the insertion section.Olympus will continue to monitor field performance for this device.
|