ABBOTT VASCULAR PRESSUREWIRE¿ X GUIDEWIRE; CATHETER TIP PRESSURE TRANSDUCER Back to Search Results
Catalog Number C12059
Device Problems Break (1069); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/16/2024
Event Type  malfunction  
Manufacturer Narrative
Visual, functional, and electrical analysis were performed on the returned device.The reported event of solid yellow light could be confirmed.Multiple bends/kinks and separation were noted on the proximal tube; however, the wire was still held together by the corewire.An electrical testing was performed, and an open circuit on the male connector area was observed during testing which could contribute to the reported issue.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no indication of a lot specific product quality issue.The investigation determined that the reported solid yellow light was likely due to the noted separation.It was likely that an excessive force was used during preparation which may have caused bends/kinks resulting in the noted separation.The separation likely caused the open circuit which could trigger the reported issue during calibration.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that during the procedure, the pressurewire x wireless device was to be used in the left anterior descending (lad) artery.The device could not be identified or connected.The transmitter light remained steady yellow.A new pressurewire x wireless device was used successfully for the procedure.There were no reported adverse patient effects and no clinically significant delays in the procedure.Return device analysis found that the proximal tube of the pressurewire x wireless device was noted to be broken at a segment that enters the patient anatomy.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSUREWIRE¿ X GUIDEWIRE
Type of Device
CATHETER TIP PRESSURE TRANSDUCER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18870765
MDR Text Key337302011
Report Number2024168-2024-03052
Device Sequence Number1
Product Code DXO
UDI-Device Identifier05415067025715
UDI-Public05415067025715
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K180558
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC12059
Device Lot Number30320G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/20/2024
Initial Date FDA Received03/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-