• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS Back to Search Results
Model Number DB-3128-55
Device Problems Fracture (1260); High impedance (1291)
Patient Problem Undesired Nerve Stimulation (1980)
Event Date 02/15/2024
Event Type  Injury  
Manufacturer Narrative
Block b3: approximated based on the date the manufacturer became aware of the event.
 
Event Description
It was reported that the deep brain stimulation (dbs) patient experienced high impedances on both sides of the leads and a loss of therapeutic stimulation.The leads and lead extensions were tested intra-operatively and it was identified that one of the lead extensions was faulty.The physician explanted and replaced the faulty lead extension.Impedances were re-checked and observed to be within range.Upon explanting the lead extension, the lead snapped.Part of the lead extension remains in-situ.The patient was doing well postoperatively and was fully recovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key18871458
MDR Text Key337290612
Report Number3006630150-2024-01361
Device Sequence Number1
Product Code NHL
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/23/2024
Device Model NumberDB-3128-55
Device Catalogue NumberDB-3128-55
Device Lot Number5000879
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/15/2024
Initial Date FDA Received03/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-