A review of the dhr and inspection records for the lot could not be conducted since a lot number was not provided.One opened sample was returned for review.Visual inspection confirmed breakage just below the inverted triangle underneath the silicone disc.Inspection under the microscope confirmed jagged edges which is indicative of a tensile force being applied to the catheter.Based on the review of the returned opened sample, the most probable cause for the breakage was most likely due to an event within the user environment.Possibly an excessive force was applied to the catheter that resulted in the breakage.The customer stated the parent transferred the baby.Since the breakage was most likely related to the user environment and not a manufacturing error, no corrective action will be taken at this time.Additional information provided in h.3., h.6.And h.9.
|