• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARGON MEDICAL DEVICES FIRST PICC S/L 26GA (1.9F) 0.65MM X 50CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARGON MEDICAL DEVICES FIRST PICC S/L 26GA (1.9F) 0.65MM X 50CM Back to Search Results
Model Number 384232
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The product was returned for analysis.Investigation including root cause analysis is in progress.A supplemental mdr will be submitted upon investigation completion.
 
Event Description
Reporter indicated "catheter broke during usage¿no further details".There was no patient injury.
 
Manufacturer Narrative
A review of the dhr and inspection records was conducted for the provided lot, and no similar concerns were found.One sample was returned for review.The catheter tubing was returned in two pieces, confirming the reported issue of breakage.Visual inspection found the breakage ends of the tubing appeared to be straight with no jagged edges as thought the tubing had come in contact with a sharp object.The most probable cause for the breakage of the catheter tubing was most likely related to an event within the user environment.Since the reported issue was most likely related to the user environment and not a manufacturing error, no corrective action will be taken.Argon will continue to monitor for issues of this nature in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIRST PICC S/L 26GA (1.9F) 0.65MM X 50CM
Type of Device
FIRST PICC
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX 75751
Manufacturer Contact
kelly mayer
1445 flat creek rd
athens, TX 75751
9036759321
MDR Report Key18872331
MDR Text Key337304463
Report Number0001625425-2024-00957
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00886333209750
UDI-Public00886333209750
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K972262
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number384232
Device Lot Number11478691
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/12/2024
Initial Date FDA Received03/10/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-