COMPANION MEDICAL INC INPEN MMT-105NNBLNA NOVO NORDISK BLUE; SYRINGE, PISTON Back to Search Results
Model Number MMT-105NNBLNA
Device Problems Computer Software Problem (1112); Material Too Rigid or Stiff (1544)
Patient Problem Hyperglycemia (1905)
Event Date 02/22/2024
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Information received by medtronic indicated that the customer reported that the inpen app was not recording the exact doses dialed on the inpen.Troubleshooting was performed, and it was unknown whether the issue was resolved.The current blood glucose value was 136 mg/dl.The customer reporting dose knob/dial was difficult to turn.The intended dose amount and amount where difficulty turning was experienced due to slipping.No harm requiring medical intervention was reported.It was unknown whether the customer will continue using the device and the inpen will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105NNBLNA NOVO NORDISK BLUE
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
lusine boyadzhyan
1223 world trade drive, 100
san diego, CA 92128
MDR Report Key18872917
MDR Text Key337424540
Report Number3012822846-2024-00317
Device Sequence Number1
Product Code FMF
UDI-Device Identifier000010862088000344
UDI-Public(01)000010862088000344(17)230920
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201337
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/20/2023
Device Model NumberMMT-105NNBLNA
Device Catalogue NumberMMT-105NNBLNA
Device Lot NumberB0656
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/26/2024
Initial Date FDA Received03/11/2024
Date Device Manufactured09/20/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
Patient Weight79 KG
-
-