• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA PRIMUS; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA PRIMUS; ANESTHESIA UNITS Back to Search Results
Catalog Number 8603800
Device Problems Gas Output Problem (1266); Failure to Analyze Signal (1539); Failure to Sense (1559); Failure to Deliver (2338)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2024
Event Type  malfunction  
Manufacturer Narrative
The investigation was just started.The result will be forwarded in a follow up report.H3 other text : on-going.
 
Event Description
It was reported that ventilation stopped during use.Manual ventilation was possible.No patient injury reported.
 
Manufacturer Narrative
Log file evaluation performed by the manufacturer confirms the reported behavior: the supervisor function detected a sporadic loss of the signal from the sensor that monitors the inspiratory cylinder pressure.To protect the patient from potentially hazardous output the device is designed to shut-down automatic ventilation and to post a corresponding "ventilator failure" alarm to alert the user.Manual ventilation with the built-in breathing bag remains possible; the monitoring functions are still available as well.Dräger finally concludes that the workstation responded as designed upon the deviation.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
 
Event Description
It was reported that ventilation stopped during use.Manual ventilation was possible.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMUS
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
Manufacturer Contact
moislinger allee 53-55
lübeck 23542
4518822868
MDR Report Key18873351
MDR Text Key337513608
Report Number9611500-2024-00104
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K042607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8603800
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/26/2024
Initial Date FDA Received03/11/2024
Supplement Dates Manufacturer Received04/16/2024
Supplement Dates FDA Received04/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/31/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-