Model Number 866199
Device Problems Display or Visual Feedback Problem (1184); Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/19/2024
Event Type  malfunction  
Event Description
It was reported to philips that the efficia dfm100 exhibited display noise.There was no reported patient involvement.The efficia dfm100 defibrillator, model# 866199, is substantially similar to the heartstart xl+ defibrillator (model # 861290) and will be reported in the united states under device model # 861290.
 
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Manufacturer Narrative
This report is based on information provided by a philips call center personnel and field service engineer (fse) and has been investigated by the philips complaint handling team.Philips received a complaint on the dfm100 defibrillator indicating a display issue, specifically that there was noise.The event was outside of use and there was no reported patient nor user harm.Available details indicate that the display was experiencing an issue.The device was evaluated onsite, the cables and connections to the display assembly were disconnected and reconnected.Upon further fse inspection, the issue could not be confirmed, however the power cable was loosely connected.The fse reseated the connections.Operational check and functional tests were performed, there were no issues.No parts were replace and no services performed.The device is fully functional and remains at the customer site.Based on the information available and the testing conducted we were unable to replicate the reported problem.The reported problem was not confirmed.The device was confirmed to be operating per specifications and no failure was identified.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EFFICIA DFM100
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC.
no.2 keji north 3rd road
nanshan district
shenzhen
CH 
Manufacturer (Section G)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC.
no.2 keji north 3rd road
nanshan district
shenzhen
CH  
Manufacturer Contact
feng she
no.2 keji north 3rd road
nanshan district
shenzhen 
CH  
7552698099
MDR Report Key18873612
MDR Text Key337334057
Report Number3030677-2024-00886
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K110825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866199
Device Catalogue Number866199
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/19/2024
Initial Date FDA Received03/11/2024
Supplement Dates Manufacturer Received02/19/2024
Supplement Dates FDA Received03/28/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/28/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-