BAXTER HEALTHCARE CORPORATION EM 2400, MAIN MODULE; SYSTEM/DEVICE, PHARMACY COMPOUNDING
|
Back to Search Results |
|
Catalog Number 2400M |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/12/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the lid magnet of a em 2400, main module door fell off.This occurred during an unspecified process step in the pharmacy.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.A visual inspection was performed, and it was noted that a magnet was missing from the lid/cover.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The missing lid magnet was identified during visual inspection.The cause of the condition was not determined.To correct the condition, the door with the magnets was replaced.The device was serviced to meet functional specification.Should additional relevant information become available, a supplemental report will be submitted.
|
|