Additional information was added to h3, h4, h6, and h10: h4: the lot was manufactured on january 2022.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation; additionally, retention samples were evaluated at the plant.A visual inspection of the photo sample was performed, and a missing cap from the set was observed.Visual inspection of the retained samples were performed with no issues noted; the samples were conforming per product specifications.The retained samples were gravity and leak tested under water; no leak was observed.The reported condition was verified on the actual sample; however, not verified on the retained samples.The cause of the condition could not be determined.The retained samples were found to be conforming product.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|