Catalog Number EMC0349
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2024
Event Type  malfunction  
Event Description
It was reported that the protective cap of an administration set for blood/derivatives was missing.This was noticed before patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
E1: initial reporter postal code: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Additional information was added to h3, h4, h6, and h10: h4: the lot was manufactured on january 2022.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation; additionally, retention samples were evaluated at the plant.A visual inspection of the photo sample was performed, and a missing cap from the set was observed.Visual inspection of the retained samples were performed with no issues noted; the samples were conforming per product specifications.The retained samples were gravity and leak tested under water; no leak was observed.The reported condition was verified on the actual sample; however, not verified on the retained samples.The cause of the condition could not be determined.The retained samples were found to be conforming product.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOOD RECIPIENT SET
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - TUNISIA
route de chebbaou
2021oued e
tunis
TS  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18875926
MDR Text Key337339197
Report Number1416980-2024-01046
Device Sequence Number1
Product Code BRZ
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberEMC0349
Device Lot Number22A24T113
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/13/2024
Initial Date FDA Received03/11/2024
Supplement Dates Manufacturer Received04/09/2024
Supplement Dates FDA Received04/15/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-