Model Number CYF-V2 |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/06/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that, during reprocessing, the videoscope tested positive for 1 colony forming unit of staphylococcus hominis.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
|
Manufacturer Narrative
|
The suspect device has not yet been returned to olympus for evaluation and the investigation is ongoing.The physical device evaluation has not been completed.Prior to the device evaluation, the device was sent out for additional microbiological testing.The microbiological analysis results are pending.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacture's (lm) review of the customer cleaning disinfection and sterilization (cds) processes, the device evaluation and the legal manufacturer's final investigation.Olympus provided the following result of the culture test, performed at the third-party labs: sampling date: (b)(6) 2024.Sampling from: all channels.Cfu: 2 cfu.Bacteria identification: bacillaceae.Sampling date: (b)(6) 2024.Sampling from: all channels.Cfu: <1cfu.The lm reviewed the customer provided the cds processes where no obvious deviations from instructions for use (ifu) were identified.After checking the contents of the cyf-v2 instruction manual, the reprocessing method is described in the following section.Chapter 6 compatible reprocessing methods and chemical agents.Chapter 7 cleaning, disinfection, and sterilization procedures.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|