This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.New information added to the following fields: b5, h3 and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over seven (7) years since the subject device was manufactured.Based on the results of the investigation, and since the device malfunction was not confirmed during evaluation, the definitive root cause of the reported issue could not be determined.The event can be prevented by following the instructions for use which state: do not coil the insertion tube, universal cord, or video cable with a diameter of less than 10 cm.Equipment damage may result.Do not attempt to bend or twist the endoscope¿s insertion section with excessive force.The insertion section may be damaged.Do not apply shock to the distal end of the insertion section, particularly the objective lens surface at the distal end.Visual abnormalities may result.If the endoscope is dropped or the distal end of the endoscope receives a hard impact, the endoscope may be damaged even if no visible damage of the lens on the distal end can be found.In this case, stop using the endoscope, and contact olympus.Do not twist or bend the bending section with your hands.Equipment damage may result.Do not squeeze the bending section forcefully.The covering of the bending section may stretch or break and cause water leaks.Do not put or press the video connector and light guide connector on the insertion section when transporting or reprocessing.The insertion section may be damaged.Olympus will continue to monitor field performance for this device.
|