MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Model Number AFAPRO28 |
Device Problems
Device Contamination with Body Fluid (2317); Protective Measures Problem (3015)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/05/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: product id: 106a3 product type: console medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during a cryo ablation procedure, a system notice was received multiple times indicating that there was a problem with the injection process.The coaxial umbilical cable was replaced and the case continued. the case was completed with cryo. it was then reported that an unknown system notice was received, while the balloon catheter was still in the patient, indicating there was blood in the catheter.The coaxial umbilical cable was unplugged and blood was observed in the coaxial port.No blood was observed in the blood bottle.Later review of the data files confirmed blood reached the coaxial port, system notices were received indicating that the safety system detected fluid in the catheter and stopped the injection, and system notices were received indicating that the safety system detected blood in the catheter handle, the injection was stopped and the vacuum disabled. no patient complicat ions have been reported as a result of this event.
|
|
Manufacturer Narrative
|
This regulatory report is a duplicate to regulatory report #3002648230-2024-00103.Any additional/new information and product analysis summaries will be sent in connection with regulatory report #3002648230-2024-00103.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|