Model Number CLH-SC
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2024
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and the evaluation found that the device using halogen light powered up for over 12 hours over 2 days.It was recommended to replace the non-olympus halogen lamp and missing spare lamp made by olympus to prevent problems from happening again.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was observed during the device evaluation, that the light source exhibited that the ceramic lamp socket unit was too loose when connected to halogen lamp.There were no reports of patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.New information added to the following fields: a review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over thirteen (13) years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause for ''lamp keeps burning up'' could not be determined.However, it is likely that the reported issue occurred due to a loose ceramic lamp socket unit.Three attempts were made to obtain additional information regarding the reported event, but no response was received from the customer.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALOGEN LIGHT SOURCE
Type of Device
HALOGEN LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18881929
MDR Text Key337846372
Report Number3002808148-2024-02296
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170024481
UDI-Public04953170024481
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K993041
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberCLH-SC
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/29/2024
Initial Date FDA Received03/11/2024
Supplement Dates Manufacturer Received03/28/2024
Supplement Dates FDA Received04/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/06/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-