It was reported, the 3-lumen sphincterotome v knife wire sparked and broke.The issue occurred while the doctor was performing an endoscopic retrograde cholangiopancreatography therapeutic procedure.The duct product was replaced in a timely manner and there were no reports of patient harm, injury or delay associated with the event.
|
Correction: d9.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the root cause of the event was unable to be identified.The event likely occurred due to one or more of the following mechanisms: 1) the device was energized in one of these following situations.(a) the cutting wire is in point contact with tissue.Or they were being close to each other.(b) the cutting wire is in point contact with distal end of endoscope.Or they were being close to each other.2) an electrical discharge occurred in the cutting wire, and the cutting wire became hot instantly.3) the cutting wire was broken.The event can be detected/prevented by following the instructions for use which state: "since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.¿be sure that the rear end of the cutting wire is extended from the distal end of the endoscope.In case the cutting wire contacts the forceps elevator, insufficient output or unintended tissue injury may occur.¿ do not activate output while the cutting wire touches the metal parts of the endoscope, or they are being close together.This could burn the tissue and/or damage the endoscope or the instrument." olympus will continue to monitor field performance for this device.
|