Model Number 471405-06
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/15/2024
Event Type  malfunction  
Manufacturer Narrative
An investigation is in progress to determine the cause of this reported event.An return material authorization (rma) was issued to evaluate the intuitive surgical, inc.(isi) device.Additional information is being gathered to determine the contribution of the device to the customer reported issue.
 
Event Description
It was reported that during a da vinci-assisted benign hysterectomy surgical procedure, the force bipolar instrument looked different at hinge tip.When the customer tried to take it out, it was unable to come out of the trocar.The customer had to use the emergency stop and undock arm to remove everything.The instrument side hinge was broken.The procedure was completed with no reported injury.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: the force bipolar instrument was inspected prior to use, and nothing was noted.When the customer was tried to remove the instrument, it was unable to come out of the trocar.The customer had to use the emergency stop and undock to remove everything.The port incision was not increased in order to remove the instrument and cannula together.There were no fragments fell into the patient.The instrument did not collide with any other instrument or tool during procedure.The force bipolar instrument jaws were not stuck in a closed position prior to attempting to remove.
 
Manufacturer Narrative
Intuitive has received the prograsp forceps instrument associated with this complaint and completed investigations.Failure analysis investigations did not replicate nor confirm the customer reported complaint.Visual inspection did not reveal any damage.The instrument was placed and driven on an in-house system and passed the recognition and engagement tests.The instrument moved intuitively with full range of motion in all directions.The grips opened and closed properly.The grips were aligned.The instrument released energy.The instrument was fully functional.
 
Event Description
Refer to h10/h11 for follow-up information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Type of Device
FORCE BIPOLAR
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key18883407
MDR Text Key337449530
Report Number2955842-2024-12328
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K214095
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number471405-06
Device Catalogue Number471405
Device Lot NumberK12230226 0077
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/15/2024
Initial Date FDA Received03/12/2024
Supplement Dates Manufacturer Received04/16/2024
Supplement Dates FDA Received05/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/23/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient SexFemale
-
-