• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND VELYS SAW INTERFACE RIGHT; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND VELYS SAW INTERFACE RIGHT; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 451570106
Device Problem Device Slipped (1584)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/07/2024
Event Type  malfunction  
Event Description
It was reported that following a total knee arthroplasty surgical procedure, it was observed that the two (x2) robotic assisted saw interfaces one right and one left were inspected, and it was confirmed that the levers of the movable parts were loose and there was a problem with fixation.Since this was not an event that occurred during surgery, there was no impact on the patient.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.This is report 1 of 2 for (b)(4).
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the evaluation has been completed, a supplemental medwatch report will be sent accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the actual device was returned for evaluation.It was found that the overall appearance of the device showed signs of normal wear from multiple uses in a clinical setting.It was found that the interaction between the sasi clamping mechanisms and saw handpiece while assembled had slight mechanical play/looseness in the direction parallel with the saw blade.The overall complaint was confirmed as the observed condition of the device would be expected to contribute to the complained device issue.The issue related to the looseness is being addressed through a capa.The assignable root cause was due to design.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELYS SAW INTERFACE RIGHT
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
DEPUY IRELAND
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic dr.
warsaw IN 46581
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18884832
MDR Text Key337416722
Report Number1818910-2024-05555
Device Sequence Number1
Product Code OLO
UDI-Device Identifier10603295519478
UDI-Public10603295519478
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K202769
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number451570106
Device Lot NumberJ45653
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/07/2024
Initial Date FDA Received03/12/2024
Supplement Dates Manufacturer Received04/16/2024
05/16/2024
Supplement Dates FDA Received04/17/2024
05/17/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/02/2021
Type of Device Usage Unknown
Patient Sequence Number1
-
-