• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 37800
Device Problems Material Deformation (2976); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Section d information references the main component of the system and other applicable components are: product id: 4351-35, lot# serial#: (b)(6), product type: lead, product id: 4351-35, lot# serial#: (b)(6), product type: lead, section d information references the main component of the system.Other relevant device(s) are: product id: 4351-35, serial/lot #: (b)(6), ubd: 25-may-2019, udi#: (b)(4); product id: 4351-35, serial/lot #: (b)(6), ubd: 25-may-2019, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider regarding a patient who was implanted with an implantable neurostimulator (ins) for gastrointestinal/ pelvic floor.It was reported that the hcp stated an x-ray was done on the pt on july 21st and it was discovered that one of the patient's lead is kinked.Hcp notes they saw a message come up when first attempting to interrogate patient's ins with 8840: 'therapy measurement is out of range.If performing an implant check the lead insertion into the neurostimulator'.Tss had the caller press 'continue' and a further impedance check was done and it was noted multiple contacts showing >20k ohms but one contact pair showed viable - c/2 at 510 ohms.Tss had the caller switch programming to this contact pair (c+2-) and the pt immediately noted they felt simulation.Caller then ran a therapy impedance measurement and it showed in-range values (10v, 22ma, 44x ohms).Tss advised they set stimulation to a comfortable level and schedule a time with the rep asap.Tss redirected to nas and provided nas number.No further complications were reported at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18885571
MDR Text Key337647638
Report Number3004209178-2024-06865
Device Sequence Number1
Product Code LNQ
UDI-Device Identifier00643169614246
UDI-Public00643169614246
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/28/2018
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/11/2024
Initial Date FDA Received03/12/2024
Date Device Manufactured06/30/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
"SEE H11...."
Patient SexFemale
-
-