SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA III XENON LIGHT SOURCE Back to Search Results
Model Number CLV-190
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/26/2024
Event Type  malfunction  
Manufacturer Narrative
In speaking with the olympus technical assistance center (tac), the customer reported that they have one tower.All scopes would not physically go into the socket correctly.The tac asked if they could see anything blocking socket holes and the customer stated no.On site support was requested but the tac explained that there are no settings to change that would affect this, and if the socket is physically broken then this would need to be repaired at olympus as these units are not on site repaired.To date, the device has not been returned.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported, the xenon light source would not go into the socket correctly.The customer needed help to attach the scope and set it up.The issue occurred after an unspecified procedure.There were no reports of patient harm.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Additional information added to fields d8, d9, h3 and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over eleven (11) years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause was not identified; however, the probable cause of the malfunction would likely due defective socket because the locking mechanism and scope detection slide were not working.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18886724
MDR Text Key337794231
Report Number3002808148-2024-02339
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/26/2024
Initial Date FDA Received03/12/2024
Supplement Dates Manufacturer Received04/09/2024
Supplement Dates FDA Received04/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/28/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-