It was reported that when opened into basket or flower, the catheter splines were tangled/intertwined.There was also an issue with irrigation.During a pulmonary vein isolation (pvi) ablation procedure, a farawave ablation catheter was selected for use.When deployed into basket or flower, the splines became tangled/intertwined.In addition, when the catheter was open, an occlusion alarm alerted on the flushing pump and the device could not be flushed.The device was manipulated in an attempt to untangle the splines but was unsuccessful.Upon successful withdrawal from the patient, there was visible spline inversion, which manually un-inverted outside of the patient anatomy.The sheath was straight during retraction into the sheath.The catheter had not been deployed without a guidewire inserted and it was confirmed that the guidewire was advanced past the catheter during deployment/undeployment.A non-boston scientific extra stiff 3mm j tip wire was utilized.The farawave did not become entangled with any other catheters during the procedure.It was noted that the patient did have small atrial anatomy.The catheter had been pulled back from the tissue when rotating the array between ablations.The catheter was replaced, and the procedure was completed successfully without patient complications.The device has been deemed contaminated by the facility and will not be returned for analysis.
|