It was reported this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.An xtw clip was prepared per the instructions for use (ifu) and inserted without issues.However, while in the valve, it was observed that one gripper was not functioning as intended.Troubleshooting was performed, but the issue was unable to be resolved.Therefore, the clip was removed and replaced.Upon removal of the clip, tissue was observed on the gripper.One clip was then deployed, reducing mr to a grade of 2.There was no clinically significant delay in the procedure.
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.A review of the complaint history identified no similar complaints from the lot.All information was investigated, and without the device to analyze, a cause for the single gripper actuation issue (during procedure) could not be determined.Additionally, a cause for the unspecified tissue injury also was unable to be determined.Tissue damage is listed in the instructions for use (ifu) as a known possible complication associated with mitraclip procedures.Serious injury/illness/impairment was a result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.
|