Model Number UHI-4
Device Problem Free or Unrestricted Flow (2945)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
H9 correction and removal number: 3002808148-10/09/2023-001-c.The device was returned to olympus for evaluation and the customer's allegation was not confirmed.Based on the results of the investigation, a definitive root cause cannot be identified.A device history review revealed no issues that could have caused or contributed to the reported issue.Device was manufactured over 4 years ago.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported, the high flow insufflation unit keeps pumping/running carbon dioxide (co2) while it is not actually being pressed/used.There was no patient harm associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18888035
MDR Text Key337794083
Report Number3002808148-2024-02353
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Repair
Type of Report Initial
Report Date 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/23/2023
Initial Date FDA Received03/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/07/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberSEE H10 FOR C&R NUMBER.
Patient Sequence Number1
-
-