Model Number A22001A
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was observed that during the device evaluation, the rigid scope's objective lens was damaged.There were no reports of patient involvement.
 
Manufacturer Narrative
The device was returned and the evaluation found that no additional reportable malfunctions other than those mentioned in b5.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely the reported failure occurred due to the effect of an excessive mechanical force caused by an impact or fall.However, a specific root cause of the reported event could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 12°, 4 MM
Type of Device
HYSTEROSCOPE, GYNECOLOGY
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg/de/germany 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg/de/germany
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18888316
MDR Text Key337479615
Report Number9610773-2024-00690
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761020848
UDI-Public04042761020848
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K897003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberA22001A
Device Lot Number638754
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/29/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/04/2024
Initial Date FDA Received03/12/2024
Supplement Dates Manufacturer Received05/15/2024
Supplement Dates FDA Received05/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/03/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-