The bronchovideoscope was cultured, but there was no growth.The scope was not ethylene oxide (eto) sterilized.Non-olympus detergents were used on the endoscope.The minimum effective concentration is being checked after each cycle in the automated endoscope reprocessor (aer).The aer used is non-olympus.The endoscope channel is being brushed during manual cleaning with a single use brush.Pre-cleaning is performed immediately after a procedure.The endoscope is stored in a non-olympus endoscope drying cabinet.An olympus endoscopy support specialist (ess) visited the customer site on (b)(6) 2024.Based on the observation summary report from the ess, there were no reprocessing deviations observed during the on-site visit.The ess provided a reprocessing in-service or reprocessing training to the user facility staff.The ess provided the user facility staff reprocessing training materials for their reference.To date, the device has not been returned.Should additional relevant information become available, a supplemental report will be submitted.This report is related to mfr's (b)(4).
|