• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRAEGER MEDICAL SYSTEMS, INC INFINITY ACUTE CARE SYSTEM (M540); PHYSIOLOGIC MONITORING SYSTEM, SINGLE-PATIENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRAEGER MEDICAL SYSTEMS, INC INFINITY ACUTE CARE SYSTEM (M540); PHYSIOLOGIC MONITORING SYSTEM, SINGLE-PATIENT Back to Search Results
Catalog Number MS25520
Device Problem Low Readings (2460)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Additional information was requested, however; a response was not received.It was not stated upon initial reporting that the patient experienced a serious injury or required medical intervention as a result of the reported issue.No device was returned to the manufacturer for physical evaluation or logs for visual inspection.Without the device or information, it is not possible to perform root cause analysis.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event Description
The customer reported that the m540 was not correctly reading an infant's o2 saturations.The m540 showed that the infant had a o2 saturation of 37%, however; the patient was awake and conscious, with a normal heart rate and not showing low o2 symptoms.The customer replaced the o2 saturation probe, and the reading remained the same.The biomed stated that there is no event/patient information as the devices were received on their desk with no additional information.There is no information stating there is any adverse patient impact or delay in treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY ACUTE CARE SYSTEM (M540)
Type of Device
PHYSIOLOGIC MONITORING SYSTEM, SINGLE-PATIENT
Manufacturer (Section D)
DRAEGER MEDICAL SYSTEMS, INC
3135 quarry road
telford PA
Manufacturer (Section G)
DRAEGER MEDICAL SYSTEMS, INC
3135 quarry road
telford PA
Manufacturer Contact
3135 quarry road
telford, PA 
9784828529
MDR Report Key18889192
MDR Text Key337482483
Report Number1220063-2024-00028
Device Sequence Number1
Product Code MHX
UDI-Device Identifier04049098009799
UDI-Public(01)04049098009799(11)210907(93)MS25520-65
Combination Product (y/n)N
PMA/PMN Number
K113798
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMS25520
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/16/2024
Initial Date FDA Received03/12/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/17/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-