Catalog Number UNKNOWN
Device Problem Excess Flow or Over-Infusion (1311)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2024
Event Type  malfunction  
Event Description
It was reported that unspecified bd infusion set had valve malfunction the following information was received by the initial reporter with the : rcc received a complaint via email.Email(s) attached.User started secondary medication (ceftriaxone) on patient through alaris pump.When looked at chamber, seemed to be running too fast (drip rate looked fast).Paused the channel medication was running through but noticed drips in chamber kept going.Clamped medication and switched medication to different channel on the same alaris pump brain.Started medication in new channel but witnessed same fast drip rate.Pump gave red alarm "channel error".Medication clamped, disconnected from patient, and removed from pump.¿ i had a video call with (b)(6) this afternoon and was able to confirm that this is a check valve failure.Took a picture of the primary drip chamber which was full.The medication is yellow and you can see discoloration in the primary bag.
 
Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Manufacturer Narrative
No product or photo was returned by the customer.The customer complaint that there is a check valve failure could not be verified due to the product not being returned for failure investigation.A device history record review could not be performed because a model and lot number was not provided by the customer.Due to no sample being received, an investigation could not be performed and a root cause could not be determined.
 
Event Description
No additional info.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNSPECIFIED BD INFUSION SET
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18891869
MDR Text Key337511251
Report Number9616066-2024-00417
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/22/2024
Initial Date FDA Received03/13/2024
Supplement Dates Manufacturer Received05/06/2024
Supplement Dates FDA Received05/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-