• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ISOFLEX OPTIM LEAD; PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ISOFLEX OPTIM LEAD; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 1944/52
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/22/2024
Event Type  malfunction  
Event Description
It was reported that the patient presented for an implant procedure.During the procedure, the atrial lead was unable to implant.The lead was not used and replaced.The patient was stable throughout the procedure.
 
Manufacturer Narrative
The reported event was the lead was unable to implant.Final analysis found that as received, a complete lead with stylet inserted was returned in one piece.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual inspection and x-ray examination of the lead did not find any anomalies.All tines were intact and undamaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISOFLEX OPTIM LEAD
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18894215
MDR Text Key337545823
Report Number2017865-2024-34848
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734502320
UDI-Public05414734502320
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
P960030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2024
Device Model Number1944/52
Device Lot NumberP000125861
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/22/2024
Initial Date FDA Received03/13/2024
Supplement Dates Manufacturer Received03/26/2024
Supplement Dates FDA Received03/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/13/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-