Model Number CLV-190
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned and the evaluation found the event previously reported in section b5 was due to a broken front panel.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was observed that during the device evaluation, the light source was not working.There were no reports of patient involvement.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Additional information added to field b5 and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.A definitive root cause was not identified, however based on the results of the investigation, the probable cause of the malfunction would likely be due to damage in the front panel.Three attempts were made to the user facility for additional information without a reply.Olympus will continue to monitor field performance for this device.
 
Event Description
The issue of the light source not working was found during preparation for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18894219
MDR Text Key337545804
Report Number3002808148-2024-02376
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/29/2024
Initial Date FDA Received03/13/2024
Supplement Dates Manufacturer Received03/19/2024
Supplement Dates FDA Received03/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/17/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-