Model Number A22003A |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was observed that during the device evaluation, the telescope, 70°, 4 mm exhibited a misaligned objective lens.There were no reports of patient involvement.
|
|
Manufacturer Narrative
|
E1.Full establishment name: (b)(6).The device was returned, and the evaluation found no additional reportable malfunctions other than those mentioned in b5.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the approved final investigation.Updated fields: h4, and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the misalignment of the objective lens was caused by improper handling and application of excessive force like fall, shock or similar stress.However, a definitive root cause could not be determined.Olympus will continue to monitor field performance for this device.
|
|