Model Number 201-30300
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/02/2024
Event Type  malfunction  
Event Description
It was reported that the centrimag system was running as left ventricular assist device (lvad) patient was in an intensive care unit bed.The pump speed was set at 3500 rotations per minute and 5-6 liters per minute.The pump speed suddenly dropped to 2900 rotations per minute and no flow.The console was exchanged, and the pump began running fine.Related regulatory report mfr # (b)(4).
 
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG 2ND GENERATION PRIMARY CONSOLE
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key18895137
MDR Text Key337560165
Report Number3003306248-2024-00450
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
K131179
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number201-30300
Device Lot NumberL06306
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/04/2024
Initial Date FDA Received03/13/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-