Model Number OER-ELITE |
Device Problem
Failure to Disinfect (1175)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported, the endoscope reprocessor had acecide-c disinfectant that was not lasting 30 days.The customer was advised that the acecide-c is a 5-day solution and not a 30-day solution.The customer was also advised that the disinfectant may need to be changed sooner than 5 days if it doesn't pass efficacy at any point.Issue addressed by educating the customer on the duration life of the acecide-c disinfectant.
|
|
Manufacturer Narrative
|
The device was not returned.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, root cause could not be identified although it can be presumed that it was caused by the user misunderstanding disinfectant replacement period: 5 days as detergent replacement period: 30 days, leading to the suggested event.The events can be detected/prevented in accordance with the following instructions for use: oer-elite operation manual.7.4 mix lcg this function mixes the disinfectant solution to enable an accurate concentration check.Note: if you attempt to perform the mrc check after a long time (maximum 5 days) since the previous use of the disinfectant, the touch screen may display the guidance for the request of the mix lcg.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|