• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. ENDOSCOPE REPROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-ELITE
Device Problem Failure to Disinfect (1175)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported, the endoscope reprocessor had acecide-c disinfectant that was not lasting 30 days.The customer was advised that the acecide-c is a 5-day solution and not a 30-day solution.The customer was also advised that the disinfectant may need to be changed sooner than 5 days if it doesn't pass efficacy at any point.Issue addressed by educating the customer on the duration life of the acecide-c disinfectant.
 
Manufacturer Narrative
The device was not returned.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, root cause could not be identified although it can be presumed that it was caused by the user misunderstanding disinfectant replacement period: 5 days as detergent replacement period: 30 days, leading to the suggested event.The events can be detected/prevented in accordance with the following instructions for use: oer-elite operation manual.7.4 mix lcg this function mixes the disinfectant solution to enable an accurate concentration check.Note: if you attempt to perform the mrc check after a long time (maximum 5 days) since the previous use of the disinfectant, the touch screen may display the guidance for the request of the mix lcg.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18896869
MDR Text Key337624460
Report Number9610595-2024-05356
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170404047
UDI-Public04953170404047
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201920
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-ELITE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/26/2024
Initial Date FDA Received03/13/2024
Supplement Dates Manufacturer Received05/13/2024
Supplement Dates FDA Received05/16/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/24/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-