INTUITIVE SURGICAL, INC ENDOWRIST; FORCE BIPOLAR Back to Search Results
Model Number 471405-06
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2024
Event Type  malfunction  
Manufacturer Narrative
Intuitive surgical, inc.(isi) has received the instrument for evaluation.However, the failure analysis has not been completed yet.A follow-up mdr will be submitted if additional information is obtained.
 
Event Description
It was reported that during a da vinci-assisted hysterectomy surgical procedure, the force bipolar instrument was unable to remove from cannula due to part of instrument catching the distal lip of cannula.The cannula removed from patient with instrument intact and pin gauge passed through cannula after removal of instrument.The procedure was completed with no reported injury.
 
Manufacturer Narrative
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The force bipolar instrument was analyzed and found to have hairline cracks on grip input disk #7.This caused inability to remove the instrument from cannula.The complaint was confirmed by failure analysis.
 
Event Description
Refer to h10/h11 for follow-up information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Type of Device
FORCE BIPOLAR
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key18899360
MDR Text Key337641223
Report Number2955842-2024-12369
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874120767
UDI-Public(01)00886874120767(10)K11230302
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K214095
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number471405-06
Device Catalogue Number471405
Device Lot NumberK11230302 0199
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2024
Initial Date Manufacturer Received 02/15/2024
Initial Date FDA Received03/13/2024
Supplement Dates Manufacturer Received05/17/2024
Supplement Dates FDA Received06/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient SexFemale
-
-