Model Number 471405-06 |
Device Problem
Difficult to Remove (1528)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/19/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Intuitive surgical, inc.(isi) has received the instrument for evaluation.However, the failure analysis has not been completed yet.A follow-up mdr will be submitted if additional information is obtained.
|
|
Event Description
|
It was reported that during a da vinci-assisted hysterectomy surgical procedure, the force bipolar instrument was unable to remove from cannula due to part of instrument catching the distal lip of cannula.The cannula removed from patient with instrument intact and pin gauge passed through cannula after removal of instrument.The procedure was completed with no reported injury.
|
|
Manufacturer Narrative
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The force bipolar instrument was analyzed and found to have hairline cracks on grip input disk #7.This caused inability to remove the instrument from cannula.The complaint was confirmed by failure analysis.
|
|
Event Description
|
Refer to h10/h11 for follow-up information.
|
|