It was reported the patient presented in clinic for follow-up.Upon interrogation, it was discovered the leadless pacemaker (lp) exhibited decreased pacing impedance, decreased p-wave amplitude sensing, and increased capture thresholds that resulting in loss of capture.Programming changes were implemented.The patient was in stable condition.
|
The reported event of low impedance, sensing problem, and failure to capture were not confirmed.The leadless pacemaker was returned for analysis.Visual inspection of the device found no anomaly on outer and inner helix, helix lengths were within specification.Further analysis performed found no anomalies contributing to reported event of impedance, sensing, or capture problem.Longevity assessment was performed, and device was in the normal range of operation with appropriate remaining longevity.
|