• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS DMC GMBH DIGITALDIAGNOST 3.0.X; SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS DMC GMBH DIGITALDIAGNOST 3.0.X; SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER) Back to Search Results
Model Number 712025
Device Problems Fitting Problem (2183); Device Fell (4014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Ref.Id: (b)(4).The digitaldiagnost r3.X is a stationary x-ray system for general radiographic purposes.As an option, a portable digital flat panel detector (model "wpd") can be used for image capture.Philips field service engineer investigated on site.He adjusted the spring mechanism on the charging station and adjusted the detecting mechanism on wall stand.After this he successfully completed gain and pixel calibrations.He tested the image and returned the system back to the customer with full functionality.Finally, the system meets meets the specification for the performed service and is returned to use.No details were provided about the circumstances when the detector fell down.As the customer did not make other claims, it is concluded that the detector was dropped by accident (use error).Risk estimation revealed acceptable risk per risk benefit analysis, because actually, there is no feasible technical solution for this kind of ¿use error¿.This issue is further monitored and trended.
 
Event Description
The customer reported that the portable wpd detector was accidently dropped and now not connecting and not charging in the docking station.A dropped detector can lead in worst case to a fracture in the foot.No injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIGITALDIAGNOST 3.0.X
Type of Device
SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS DMC GMBH
roentgenstrasse 24
hamburg 22335
GM  22335
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS DMC GMBH
roentgenstrasse 24
hamburg 22335
GM   22335
Manufacturer Contact
dusty leppert
100 park avenue, suite 300
orange village, OH 44122
MDR Report Key18902711
MDR Text Key337669660
Report Number3003768251-2024-00008
Device Sequence Number1
Product Code MQB
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K131483
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number712025
Device Catalogue Number712025
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/15/2024
Initial Date FDA Received03/14/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-