• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 7122Q/65
Device Problems Failure to Capture (1081); Fracture (1260); High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/22/2023
Event Type  Injury  
Manufacturer Narrative
Further information was requested but not received.
 
Event Description
It was reported that the patient presented for remote follow-up via merlin.Net with recorded episodes of non-sustained right ventricular oversensing.A review of the transmission revealed that the episodes were caused by loss of capture exhibited by the right ventricular lead.Also noted were gradual increases in both pacing and defibrillation lead impedance measurements.No patient symptoms or interventions were reported.Further information was requested but not received.
 
Manufacturer Narrative
Additional information: b5 and h6.
 
Event Description
New information received notes that lead fracture was suspected.
 
Event Description
New information received notes that the lead was capped and replaced on (b)(6) 2024.The patient tolerated the procedure well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18904893
MDR Text Key337688892
Report Number2017865-2024-35014
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734503341
UDI-Public05414734503341
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2013
Device Model Number7122Q/65
Device Lot Number3811314
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/23/2024
Initial Date FDA Received03/14/2024
Supplement Dates Manufacturer Received03/14/2024
03/13/2024
Supplement Dates FDA Received03/15/2024
03/29/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/16/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ELLIPSE; RIGHT ATRIAL LEAD
Patient Outcome(s) Required Intervention;
Patient SexMale
Patient Weight111 KG
-
-