• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE AUTOG BC PNK 20GA X 1.0IN; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE AUTOG BC PNK 20GA X 1.0IN; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 382533
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2023
Event Type  malfunction  
Event Description
It was reported that bd insyte autog bc pnk 20ga x 1.0in catheter contained a break.The following information was provided via medwatch: the iv needle was taken out of the packaging and the reservoir was slightly moved to loosen up.Registered nurse (rn) noticed that the tip of the catheter was almost breaking off and hanging by a thread.Rn disposed of damaged catheter.
 
Manufacturer Narrative
A.2.Date of birth: unknown.The patient¿s age was used to determine a placeholder date for this field.E4.The initial reporter also notified the fda november of 2023.Medwatch report is (b)(4).H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Manufacturer Narrative
Investigation results: a device history record review was completed by our quality engineer team for provided material number 382533 and lot number 3142661.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.Complaints received for this device and reported condition will continue to be tracked and trended.
 
Event Description
Additional information: describe any patient harm, injury, complication or negative outcome that occurred as a result of the event.No patient harm or injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE AUTOG BC PNK 20GA X 1.0IN
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18906048
MDR Text Key337689841
Report Number1710034-2024-00185
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00382903825332
UDI-Public(01)00382903825332
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number382533
Device Lot Number3142661
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/21/2024
Initial Date FDA Received03/14/2024
Supplement Dates Manufacturer Received03/21/2024
Supplement Dates FDA Received04/01/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age71 YR
-
-