It was reported that this pacemaker device is suspected of behaving in a manner consistent with a premature battery depletion (pbd).The estimated battery longevity in april 2023, was 2 years remaining.In october 2023, the remaining batter longevity was 4 months.A patient will follow up in the near future for an in-office visit.The device remains implanted.No adverse patient effects were reported.New information was received indicating the patient returned to the clinic for a follow up appointment.A request was made to have a technical service (ts) consultant review data disk analysis from this device.Data analysis confirmed the battery appeared to be depleting normally.The device remains implanted.No adverse patient effects were reported.Additional information was received indicating that this device was explanted due to a suspected premature battery depletion (pbd).A new device was implanted.Besides surgical intervention, no adverse patient effects were reported.
|