Model Number J276
Device Problems Premature Discharge of Battery (1057); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2023
Event Type  Injury  
Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event Description
It was reported that this pacemaker device is suspected of behaving in a manner consistent with a premature battery depletion (pbd).The estimated battery longevity in april 2023, was 2 years remaining.In october 2023, the remaining batter longevity was 4 months.A patient will follow up in the near future for an in-office visit.The device remains implanted.No adverse patient effects were reported.New information was received indicating the patient returned to the clinic for a follow up appointment.A request was made to have a technical service (ts) consultant review data disk analysis from this device.Data analysis confirmed the battery appeared to be depleting normally.The device remains implanted.No adverse patient effects were reported.Additional information was received indicating that this device was explanted due to a suspected premature battery depletion (pbd).A new device was implanted.Besides surgical intervention, no adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALIO
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18907277
MDR Text Key337729459
Report Number2124215-2024-15636
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeNL
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/17/2016
Device Model NumberJ276
Device Catalogue NumberJ276
Device Lot Number103732
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/08/2024
Initial Date FDA Received03/14/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/20/2014
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age86 YR
-
-